LAQUA PC1100

Share:


LAQUA PC1100

Contact Us
Izah
: +6018-236 7662
: ecc17info@gmail.com

Model: LAQUA EC1100 EC/Res/TDS/Sal/Temp (°C)
pH Range:
-2.000 to 20.000pH
  • Resolution: 0.1/0.01/0.001 pH
  • Accuracy±0.003 pH
  • Calibration points: 5
  • Buffer Options: USA, NIST, Custom
ORP Range:
±2000 mV
  • Resolution: 0.1 mV
  • Accuracy: ±0.2 mV
Conductivity (EC) Range:
0.000 µS/cm to 20.00 mS/cm (k=0.1)
0.00 µS/cm to 200.0 mS/cm (k=1)
0.0 µS/cm to 2.000 S/cm (k=10)
  • Resolution: 0.05% of full scale
  • Accuracy: ±0.6% of full scale ; ±1.5% of full scale > 18.0 mS/cm
  • Reference Temperature: 15 to 30°C (Adjustable)
  • Temperature Coeficient: 0.00 to 10.00% (Adjustable)
  • Cell Constant: 0.1 / 1.0 / 10.0 (Selectable)
  • Measurement Unit: µS/cm, mS/cm, S/m (Auto-ranging / Manual)
  • Calibration Points: Up to 4 (Auto / Manual)
TDS Range:
0.00 mg/L to 100 g/L
  • Resolution: 0.01 mg/L
  • Accuracy: ±0.1% of full scale
  • Calibration Curves: EN27888 / 442 / Linear (0.40 to 1.00) / NaCl
Resistivity Range:
0.0 Ω•cm to 2.000 MΩ•cm (k=10)
0.000 kΩ•cm to 20.00 MΩ•cm (k=1)
0.00 kΩ•cm to 200.0 MΩ•cm (k=0.1)
  • Resolution: 0.05% of full scale
  • Accuracy: ±0.6% of full scale; ±1.5% of full scale > 1.80 MΩ•cm
Salinity Range:
0.00 to 100.00 ppt / 0.000 to 10.000%
  • Resolution: 0.01 ppt / 0.001 %
  • Accuracy: ±0.2% of full scale
  • Calibration Curves: NaCl / Seawater
Temperature Range:
-30.0 to 130.0°C
  • Resolution: 0.1°C
  • Accuracy: ±0.4°C
Memory Data: 999
Auto-Hold / Auto-Stable: Yes
Electrode Status: On screen display
Diagnostic Messages: Yes
Display: Custom LCD, Dual channel display
Inputs: Dual BNC, dual phono, DC sockets
Output: RS232C, USB
Power Requirement: AC adaptor, 100 - 240V, 50/60 Hz
Weight: 500 g
Dimensions: 170(L) x 174(D) x 73(H) mm

 

Specifications

Sorry, currently do not have any product specifications

How to use LAQUA PC1100

Sorry, currently do not have any file for download.